Jon also fixed map issues and display of the scaler with the F1 TDC with Stewe Wood and looked but I guess he can be more accurate about that than I.
I took a quick pedestal run 1510 before leaving when they were done with the DAQ : F1 TDC were giving error message, I think Robert Feuerbac is looking at the problem right now and Jon might continue tomorrow work on that tomorrow so all F1 will be running.
Pending issues with neutron arm after a quick look at the pedestal we still have : s1-02-L s1-03-L n4-06-L n1-05-L n1-26L v1-39-L v1-40-L ( wide even without veto ! ) n1-15-R, n3-31-R, n4-16-R n6-34-R s1-07-R, s1-08-R with wide pedestal values to be checked.
This could be due to light leaks, amplifier or whatever... so I will turn off the HV on those channels. Don't forget to turn them back on if you look at them ! I did not check if all the HV were by the way, so we will have to take a pedestal run with everything on.
So getting there...
In attachement, the HV status after I turned off the forementionned HV channels and a pedestal display of run 1514 (looks like HV off did not improved anything ).
Regards,
Alexandre
A copy of this log entry has been emailed to: mjona@jlab.org,tajima@jlab.org,shahinya@jlab.org,bogdanw@jlab.org,camsonne@jlab.org
Neutron detector High Voltage Left
Neutron detector High Voltage Right
Pedestal
Pedestal PS file
Pedestals_1514.ps