User name R.M./B.Craver
Log entry time 12:12:59 on February25,2006
Entry number 163823
This entry is a followup to: 163813
keyword=ideas to fix bad7 problems
There is evidence chamber 1 causes the problem since we only
changed its thresholds, and a high threshold (9) greatly reduced
the noise rate which paralyzes TDC chips. So.... the ideas to
fix this problem:
1. Raise thresholds on chamber 1. And raise HV to verify the
problem (bad7's) stays gone.
2. Chamber 1 is unique in that it has new preamp cards. One might
imagine putting back the old cards.
In any case the hope is to reduce the bad7 rate with cosmics to
zero, then perhaps further fine tuning of threshold/HV during beam.
It is not obvious that TDC 1877 would not have the same problem ...
we know that low thresholds on HRS cause these to be "paralyzed",
so replacing TDC models is not sure to cure this. Perhaps better
to work on the noise.