User name Miller / Camsonne / Feuerbach
Log entry time 23:41:45 on March 01, 2006
Entry number 164693
This entry is a followup to: 163824
keyword=No Veto because of amplifier NIM crate problem
When looking at signal at the input of the discriminators we saw a 100 mV
positive DC offset and no pulses.
After power cycling the amplifier NIM crate, usual signal could be seen.
This crate alread had some power supply issues : I am not sure if the
power supply or the whole crate was changed, so if this was only the
power supply that might hint a problem with the backplane. So this has to
be investigated at the next access tomorrow.
A copy of this log entry has been emailed to: shahinya@jlab.org,bogdanw@jlab.org,mjona@jlab.org,tajima@jlab.org