    User name kelleher

    Log entry time 18:46:27 on March 4,2006

    Entry number 165201

    keyword=target polarization = 29.1%

    Performed target polarization and calibration. I had a hard time getting an EPR signal (my mistake, nothing wrong with the system), so I have no doubted that I significantly depolarized the cell when I took the measurement. However, I got a calibration number that supports Gordon's theory of how the calibration "constant" should change. It continues to increase.

    My new calibration constant is 0.5057 %/mV. Since this is higher than the previous constant and my measurement is suspect, I suggest that we continue to use the more conservative 0.465% until we can confirm this measurement.