User name tech /folts
Log entry time 13:09:51 on March 17, 2006
Entry number 167060
Followups:
keyword=replaced enterance window assembly
Replaced enterance window assembly. On inspection we could see
deformation of the Be window that appeared to be caused by heat. Pictures
of the window appear to show a hot spot in the center of the square
raster. The pix are not great quality but can be found on the Hall A tech
page.
We were not able to see a hole in the wondow and a leak check was not
done due to radiation, and possible be contamination. however when the
new assembly was pumped down and leak checked,the vacuum problem was
resolved.