I've calculated a new set of momentum calibration coefficients which
should be valid for the 2.5 and 3.4 kinematic. Below are the plots of
the difference between calculated momentum and what we expect from the
electron scattering angle by itself, against focal plane X and inverse
deflection angle. The resolution appears to be on the same level as what
we had before (around 1.5%ish). I'll try and get more statistics to get
a better idea of what our resolution is.
These coefficients will be put into the databases running from 03/09 to
present shortly.
Also, the BigBite magnet readback appears to have changed from 710 to 706
beginning March 17. I'll investigate if this has an effect on the
momentum reconstruction. Fortunately, we have some hydrogen runs taken
on the 9th or 10th which should be sufficient to do a recalibration for
this period of necessary.