User name kelleher
Log entry time 13:33:07 on March 26, 2006
Entry number 168434
keyword=NMR measurement
Took NMR measurement at 13:13, but failed to turn on the upper lockin, so
flipped the spins but I have no signal.
Took another one at 13:16, with phase set to 0 (normally 87.6) so most of
the signal is in the Y channel, X channel couldn't be fit with the
current program.
Lower:
UP = -89.55 +/- 0.58
DOWN = -88.65 +/- 0.64
Upper:
X
UP = -1.254 +/- 0.011
DOWN = -1.175 +/- 0.0094
Y
UP = 0.5504 +/- 0.022
DOWN = 0.6099 +/- 0.010
Polarization: 38.4%
So, clearly, we are polarizing again. Right now, it's hard to say if we
are polarizing at the rate that we saw before, but we will find out after
a couple more measurements. We will take measurements every 3 hours
until the polarization is back above 50%.