Main INDEX
Monthly INDEX
PREV
NEXT
User name kelleher
Log entry time 16:48:49 on April 06, 2006
Entry number 169781
keyword=Target polarization measurement
Performed an NMR-EPR calibration. We asked for beam to be taken away at
15:48, and asked for beam back at 16:01.
The measurement includes many AFP sweeps, and as a result, we have lost
some polarization due to the measurement.
Our final NMR result is listed below:
Lower Coils:
X
UP = 118.2 mV +/- 0.785
DOWN = 117.7 mV +/- 0.881
Y
UP = 3.730 mV +/- 0.079
DOWN = 3.179 mV +/- 0.087
Upper Coils:
X
UP = 1.466 mV +/- 0.009
DOWN = 1.432 mV +/- 0.012
Y
UP = 0.7316 mV +/- 0.0075
DOWN = 0.7619 mV +/- 0.013
Polarization = 50.9%