    User name Yi Qiang

    Log entry time 07:59:01 on April 18, 2006

    Entry number 171169

    keyword=Instructions for locally setting NMR RF Function Generator

    The NMR RF function generator failed twince during this owl shift, when
    that happens, you will get no response after your click the go button in
    NMR measurement. In case that occurs, here are instructions from Aidan:

    1. Set generator mode to local
    press "Local" button.

    2. Set Frequency to 91 kHz
    press "Freq" button in PARAM section, then "9", "1" and "kHz ms mV".

    3. Set Amplitude to 810 mV
    press "Ampl", then "8", "1", "0", BLUE button(shift) and "kHz ms mV".

    4. Check signal
    press "Signal on off" button above the right BNC connector, and see
    whether there is signal on NMR osciliscope. Once comfirmed, press that
    button again to turn it off.

    5. Continue NMR measurement.