User name Yi Qiang
Log entry time 07:59:01 on April 18, 2006
Entry number 171169
keyword=Instructions for locally setting NMR RF Function Generator
The NMR RF function generator failed twince during this owl shift, when
that happens, you will get no response after your click the go button in
NMR measurement. In case that occurs, here are instructions from Aidan:
1. Set generator mode to local
press "Local" button.
2. Set Frequency to 91 kHz
press "Freq" button in PARAM section, then "9", "1" and "kHz ms mV".
3. Set Amplitude to 810 mV
press "Ampl", then "8", "1", "0", BLUE button(shift) and "kHz ms mV".
4. Check signal
press "Signal on off" button above the right BNC connector, and see
whether there is signal on NMR osciliscope. Once comfirmed, press that
button again to turn it off.
5. Continue NMR measurement.