    User name A. Camsonne

    Log entry time 14:21:39 on April 21, 2006

    Entry number 171459

    keyword=Morning shift summary

    Smooth running
    efficiencies of run
    4103
    4105
    looks stable.
    No problem seen on replays either.
    Rough calibration from the elastic peak
    Detector
    Width
    in cm
    Height
    E Loss in MeV
    protons
    ADC proton Eloss
    ADC calibration
    100 channels
    in MeV
    Threshold in mV
    Threshold in ADC
    Threshold in MeV
    Veto
    2

    3.7
    300
    1.23
    15
    150
    1.8
    CMU
    10

    18.5
    400
    4.625
    20
    150
    7
    UVA
    20

    37
    1000
    3.7
    20
    150
    5.6
    Gla
    25

    46.25
    374
    12.4
    20
    150
    18.6
    This gives a threshold on the sum of around 45 MeV for -180 mV ( for 600 ADC channels )