User name M. Olson
Log entry time 19:02:57 on August 6,2006
Entry number 177760
This entry is a followup to: 177755
keyword=Loop 2 (LH2) JT adjustments, response sluggish
I adjusted the loop2 (LH2) JT valve downward (closed) by 1.5%, in 0.5% increments, to decrease the amount of overcooling in the loop.
The loop is currently set up for 20uA with the following assumptions:
Beam heating @ 20uA: 100W
Control overhead: 30W
Additional allowance for transients: 20W
Total Loop 2 heat load: 150W
Note: The observed response to these 0.5% adjustments appeared to be quite slow, on the order of 15-20 minutes overall. The response when I rebalenced the heating was much faster, almost immediate, so I am a bit confused as to what the difference was in this case.