User name A. Camsonne
Log entry time 18:26:11 on August24,2006
Entry number 182315
keyword=Trigger scintillator HV tune
We looked at the ADC specta with Bogdan, efficiency coould be immproved by better balancing of HV :
Left S1 left PMT goes from 1950 V to 1980 V
Left S0 left PMT goes from 1500 V to 1600 V
Left S0 right PMT goes from 1500 V to 1600 V too