keyword=Cycling the power on the NMR meters
Sometimes the NMR meters need to be power cycled. Most likely when the meters do not respond to GPIB commands. To check if the meter is responding one can try to change the units on the NMR GUI. It's the GUI you get to from the "FIELDS" button and selecting "D1 NMR". Try changing units from "MHz" to "Tesla". If the units on the display change, the meter is responsive and does not need to be power cycled. If the units do not change it would be good to power cycle the meter. The AC Power Switch for the Left Arm is hareboot17, port 8. The AC Power Switch for the Right Arm is hareboot27, port 8.