User name M Jones
Log entry time 06:19:39 on October 8,2006
Entry number 186577
keyword=Right arm problem
While looking for something else noticed that in run 1142 start-of-run
it wasn't able to read the right dipole. Looking more closely at the
peak in the Erecoil spectra I see that it has drifted relative to
the previous run 1141 and 1140. Also run 1143 the erecoil has
drifted.
In the xterm connected to ioc16 the message:
0x18e0890 (CA TCP): CAS: Client accept error was "S_iosLib_TOO_MANY_OPEN_FILES"
Reboot ioc16
Quads field and current values are updating in the Tools screen.
The NMS for dipole is stuck. Doesn't respon to changing the Units.
Power off/on NMR from the web interface. Now NMR seems to function
but the search is stuck in OFF.
Reboot ioc16, power off/on NMR still the dipole current doesn't respond
Reboot ioc16, power off/on NMR
Reboot ioc16