    User name M Jones

    Log entry time 06:19:39 on October 8,2006

    Entry number 186577

    keyword=Right arm problem

    While looking for something else noticed that in run 1142 start-of-run
    it wasn't able to read the right dipole. Looking more closely at the
    peak in the Erecoil spectra I see that it has drifted relative to
    the previous run 1141 and 1140. Also run 1143 the erecoil has
    drifted.
    In the xterm connected to ioc16 the message:

    0x18e0890 (CA TCP): CAS: Client accept error was "S_iosLib_TOO_MANY_OPEN_FILES"

    Reboot ioc16

    Quads field and current values are updating in the Tools screen.
    The NMS for dipole is stuck. Doesn't respon to changing the Units.
    Power off/on NMR from the web interface. Now NMR seems to function
    but the search is stuck in OFF.

    Reboot ioc16, power off/on NMR still the dipole current doesn't respond

    Reboot ioc16, power off/on NMR

    Reboot ioc16