keyword=Owl shift Summary
Recovering from right arm dipole problem when shift starts.
At 00:38 start run 1140 at A7 kinematics. Take 5M.
Run 1141 5M
Run 1142 5M
Switch to kinematics A5 with seive in the left arm
Run 1143 5M
Run 1144 4.7M
During run 1144 realize that the right dipole stopped communicating
at least since the beginning of run 1142. Looked closely and
see that the Erecoil peak had shift in run 1142 relative to 1140 and 1141.
Try reboot the ioc16 and number of times but can't regain control
of the right arm dipole current. Page Mark Stevens to come in.
In controlled access with Mark Stevens in hall.