User name R. Michaels
Log entry time 05:42:30 on October13,2006
Entry number 186979
This entry is a followup to: 186977
keyword=reboot R-HRS NMR too
Rebooted R-HRS NMR too.
This semed to be necessary because after rebooting IOC the NMR
was frozen. After rebooting NMR, we found we had to:
1. Put to Mode=Auto, Field Pol=Neg.
2. Want Probe B, but it was reading back probe A. Click on the
pulldown menu for Probe# to wake it up to read back B.
3. Wanted Search=On, but this refused to turn on until we clicked
Regulator Off. Then Search=On worked, then regulator=On.
4. Finally the NMR locked.
(As per the sticky note, "play if needed")
Then Alexandre shows up and says we didn't have to do any of
this because the R-arm dipole current is stuck (indeed it isn't
fluctuating now either).
Started run 1279.