Main INDEX
Monthly INDEX
PREV
NEXT
User name Yi Qiang
Log entry time 19:33:40 on October22,2006
Entry number 188001
keyword=Right D1 NMR control reboot
We first lost the IOC control of Right HRS Magnet. After reboot, all the other feedback is OK except D1 NMR. We therefore rebooted the Right NMR control on hareboot27. The NMR could not find the field automatically, so we had to manually set the DAC to around 2500 first and do the search again. Now the right dipole is back.