User name A. Camsonne
Log entry time 00:37:27 on October23,2006
Entry number 188034
keyword=Target OK HPH Loop 3 power during trips
The target is fine, though it seems the low limit for HPH is 450 and high limit is 500 W, these values can be excedeed briefly during a beam trip. We can reduce a bit the flow or increase the limit of the value of the High Power Heater on loop 3. When an alarm is issued the alarm stays on even if the value is back to normal so one can safely acknowledge it if the value is back to under 450 W.