User name R. Michaels
Log entry time 06:07:29 on December14,2006
Entry number 190699
keyword=raster repair
M. Olsen and I replaced the raster power supply for the vertical.
We had to get one out of room 126 in bldg 90. The one we replaced
was driver#8 and the new one is labelled #1.
I'm pretty sure the problem is the 24V supply, because I saw the
same problem in ~Oct and traced it to that supply. The LED
indicator, which goes directly across this supply, was out. Also
it is a common failure mode. So, I will buy ~5 more supplies.
The part number is SP-500-24 from astrodyne.com (800-823-8082).
Other issues were that the one that was broken in Oct did not get
fixed probably (it's still on the floor in the hall) and no spares
are in
the counting
room. We'll need to replace 24 supplies and collect
spares in
order to minimize downtime in the future.