    Shift crew: Robert Feuerbach and Kalyan Allada


    0:30 Requested controlled access to insert pin for LHRS small angle limit after moving to 14.046 deg for Kinematic 'X'.

    1:12 Started runs 6407/26384

    1:24 Noticed VDC inefficiency was approaching 10% at the elastic peak. Requested the beam current to be dropped to 0.3 microAmps for the next run. Found 0.4 microAmps is the minimum to get a readback from the BPMs and will run with it.

    2:08 Switched to B4C target.

    2:50 Switched to Carbon target.

    3:30 Switched to Ta target, and raised current to 1 microamp to be consistent with the Carbon data. Next run will be at 0.4 microamp to provide data to compare and understand the VDC/tracking efficiency. (Dedicated data for this purpose might be called for later)

    4:00 DAQ stopped collecting data for the LHRS. Rebooted ROCs/TS and restarted coda.

    6:30 Started moving to 46deg for kinematic F.

    7:05 Starting kinematic F points.

    7:30 Found raster patterns to be 1mm by 3mm (see halog entry). Moved to empty target to try to improve raster shape. Runs 6423-6425 were used to test the raster size.

    7:55 Alexandre was kind and took over for me for the last few minutes. Thanks!.





    Run Summary


    Runs Kin Target Start Comment
    6407/26384 X Li 1:07 0.5 microamp
    6408/26385 X Li 1:31 0.4 microamp
    6409/26386 X Li 1:47 0.4 microamp
    6410/26387 X B4C 2:11 0.4 microamp
    6411/26388 X B4C 2:28 0.4 microamp
    6412/26389 X C 2:51 0.4 microamp
    6413/26390 X C 3:08 0.4 microamp
    6414/26391 X Ta 3:40 1.0 microamp
    6417/26394 X Ta 4:20 0.4 microamp
    6418/26395 X Ta 4:50 0.19 microamp
    6419/26397 X C 5:16 0.19 microamp
    6421/26399 X B4C 5:49 0.19 microamp
    6422/26400 X Li 6:12 0.19 microamp

    6426/26404 F Li 7:52 4, then 3 microamp