    User name Aniol

    Log entry time 08:28:53 on March 9,2007

    Entry number 195299

    keyword=bismuth Iscan

    MCC raster numbers (8,7.4), (4mm,7mm)
    I uA T1(Hz) T3(Hz)
    5.3 1K 930
    10.5 2.1K 1.83K
    21 3.95K 3.61K
    42 8.41K 7.09K
    63.7 13.5K 10.7K
    84.6 18.5K 14.4K

    The rate on the Bi tgt is bigger than the rate on the Pb tgt.

    I(Pb)79uA : T3 = 9.16K
    I(Bi)84.6uA: T3 = 14.4K,
    rates Bi/Pb = 1.47

    We reduced the raster size by a factor of 2 and decreased the current to 20 uA to have the same power density.
    MCC raster numbers (4,3.7), (2mm,4mm) The pattern is below.

    I T1 T3
    21uA 3.95K 3.6K at (4mm,7mm) from above
    22uA 5.4K 3.9K at (2mm,4mm)
    22uA 4.8K 3.6K at (4mm,7mm) going back to normal raster

    Conclusion: Reducing the raster size slightly increases the rate ,so we are not hitting the ladder. I expected the quasielastic rates in T3 to be the same for Bi and Pb since the targets have about the same mass/cm2.

    The second raster pattern is for 85 uA run 1326, production on Bi.


    FIGURE 1

    FIGURE 2