    Swing shift workers: Makis Petratos, Milan Potocar, Mina Katramatou

    At beggining of shift rebooted IOC for right spectrometer.

    Continuing with Kinematics 02:

    16:20-16:46 1733, target C, 43 uA, 0.061 C

    Target changed to Pb+C#3

    16:54-17:05 1734, Pb+C, 10 uA, 0.006 C
    17:05-17:17, 1735, Pb+C, 20 uA, 0.012 C
    17:18-17:28, 1736, Pb+C, 30 uA, 0.016 C
    17:34-18:00, 1737, Pb+C, 42 uA, 0.061 C
    18:00-18:30, 1738, Pb+C, 42 uA, 0.061 C
    18:30-18:57, 1739, Pb+C, 40 uA, 0.061 C
    18:58 1740, Pb+C, 40 uA, CODA CRASH! No end of run!

    19:15 CODA crash: required entry downstairs to manually
    reset the Right Arm Port Server and two Fast Bus
    crates.

    During Hall entry, the power for the camera motion
    system for the Left Arm was cycled.

    20:30-20:41, 1742, Pb+C, 10 uA, 0.005 C RUN NO GOOD/CODA PROBLEMS

    21:05 ROC1 problem, need to power off and on the fastbus power supply down in the Hall, requested hall access.
    21:40 back from Hall, reqested beam at 10uA

    Runs 1743-1745 are test runs, not to be used for data

    21:48-21.59, 1746, Pb+C, 10uA, 0.006 C
    22:00-22:10, 1747, Pb+C, 20uA, 0.012 C
    22:13-22:39, 1748, Pb+C, 40uA, 0.061 C

    From analysis of run 1748 noticed missing ADC spectra of Right arm
    right S2s. Investigating this.

    1749 pedestal run
    1750 test run
    1751 pedestal run

    22:45 Requested Hall access to check electronics
    Turns out problem was due to a very flaky ribbon cable

    1752 cosmic ray run, all looks ok

    23:25 asked for beam back

    23:41-23:54, 1753, Pb+C, 10uA, 0.007 C