• Main INDEX
  • Monthly INDEX
  • PREV
  • NEXT

    User name X. Zheng

    Log entry time 10:36:25 on March25,2007

    Entry number 197396

    Followups:

    keyword=Left magnet alarm

    Around 10:05 there was an alarm on HRSL magnets. At the first all HRSL magnet fields were blank. Before I can react (reboot IOCs) they all came back. I typed in the momentum setting, quads are okay but the NMR lost lock. Setting NMR magnet 3 times made it back to 2.159 only (the correct setting is 2.2160). I tried both probe B and C. Clicking on "MHz" and "Tesla" and the GUI responds correctly (no need to power cycle according to instructions). Tried also turning the regulator off and on.

    Just in case, also tried to connect to hareboot17 for left NMR reboot but seamonkey shows "failed to connect".


    FIGURE 1

    FIGURE 2