    User name sawatzky

    Log entry time 11:17:35 on April 16, 2007

    Entry number 199259

    This entry is a followup to: 199258

    keyword=Gas panel alarm followup

    I spoke with Jack Segal. The Interlock alarm is likely due to wind gusts screwing with the gas hut exhaust fan (like Javier said).

    The fault light on the Target Gas sensor box is NOT a problem. (That circuit is not being used right now).