    We need to have the stepper voltage at 18V minimum (which is the voltage we are using now). If the voltage is lower than that the motor motion is smoother, BUT the motor gets "stuck" at the UL (radiator out) limit switch.

    Pulse Generator: 100 Hz, 100 Hz square wave 0V min, +2V max. *** Confirm the pulse on the scope. ***

    For the Stanford Model DS345 unit we have now the function generator settings to get what we need are: 100 Hz, 1V amplitude, 0.5V offset. This generates the needed 0--2V square wave on the scope. Go figure.