keyword=S0 check
I took the opportunity of the controlled access to look at S0 signal
again, both signals looks similar with still a small reflection about 150
ns after the pulse. I tried to see if the BNC cable was well seated on
both S0 PMT, they were. But I did not have time to check after the
discriminator since Heidi and Mark were done. The discriminator channel
or the TDC channel might be a bit noisy but if we cut on the right timimg
peak the ADC spectrum looks clean. I will try to move to another
TDC/discriminator channel at next pass change.