    User name Techs

    Log entry time 13:52:16 on June 24, 2007

    Entry number 206481

    keyword=Replaced middle bogie motor #4

    After a series of adventures Scot replaced the middle, outer bogie motor. The motor controller gave an over speed fault when the wheel hit a crack in the floor plate. At one point the speed fault on the next try was on motor 3, then the next try was motor 2. The spectrometer would back up and run ok, till we tried to cross the crack. Motor 4 would fault on over speed. We tried putting a couple of different shims under the wheel at the crack, to no effect. Eventually we had motor 4 spin up while 1, 2, and 3 just sat there. It did that three times in a row. We got a spare motor and swapped it out. Spectrometer started running ok.