    keyword=tech - Left Q3 work during polarity change

    Verified that the Left Q3 power supply can be turned ON and OFF remotely. It will accept a set voltage in either local or remote mode, it just doesn't ramp up. The spare SE20 module was verified to be of similar wiring to the one currently in the supply. When the spare was installed the supply responded to the reset button, but not the power on button. With the spare SE20 installed the supply could be turned on remotely and it did accept a set voltage while in remote. The supply would not ramp up. The original SE20 was re-installed. The spare SE20 was tagged and put on a shelf in the trailer.