    User name fansler

    Log entry time 15:27:58 on November 08, 2007

    Entry number 213020

    keyword=Right Q1/D1 He read back circuit

    The right Q1 He readback circuit has been converted from 2-10 V to 0-10V to the hall A tools page (1313 ch 18). The cryo readback is still 4-20 ma converted to 2-10V at cryo's cross connect rack. Both circuits are isolated with Pheonix isolation modules.

    A UPS has been installed for the Dipole quench detection and Dipole He level circuits to provide a dedicated power source for the Wang protection circuit. The lead flow circuits are still routed through the wang protection circuit, Securing power on the Wang protection Circuit UPS.... WILL FAST DUMP THE MAGNET POWER SUPPLY.