• Main INDEX
  • Monthly INDEX
  • PREV
  • NEXT

    User name V. Sulkosky

    Log entry time 15:37:42 on November21,2007

    Entry number 215979

    keyword=BPM positions after Ped Update

    Figure 1 shows the BPMA and BPMB positions after I updated the
    pedestals in the db_rb.Raster.dat, db_urb.BPMA.dat and db_urb.BPMB.dat
    database files for the CSR database directory: DB/20071108. Figure
    2 shows the BPM positions before updating the pedestals. The plots
    are for Run 2150, See Halog entry 213381

    Unfortunately, updating the pedestals made matters worse. Hence, the
    calibration constants are incorrect with data using the "new"
    attenuator module.

    To fix this, I will first calibrate the BPMs from the Bullseye scan
    using the original attenuator. I might then be able to use the
    RHRS BPM information to extract a calibration during the period of
    using the "new" attenuator on the LHRS.


    FIGURE 1

    FIGURE 2