User name R. Michaels
Log entry time 06:28:36 on December11,2007
Entry number 219873
keyword=R-HRS scaler checkout
Now to the R-HRS scalers. I looked at the same plots as for
L-HRS but don't post that. Instead I focus on the alleged problem
of BCMs. Indeed d3 looks strange but u10 looks more or less ok.
Maybe something has changed since halog 219705. See fig 1 (rates).
[adaq@adaql6 scaler]$ ./tscalntup /adaql1/data3/e05110_21748.dat.0 Right
Bank = Right
Opened scaler map file scaler.map
Bdate : day = 8 month = 10 year = 2007
Using Right Arm BCM pedestals
Normal end of file /adaql1/data3/e05110_21748.dat.0 encountered
------------- Scaler Summary ----------------
Scaler bank Right
clock location 3 0
clock counts 31585043 clock rate 103760
Time of run 5.07 min
Triggers: 1 = 41294 2 = 70231 3 = 108488 4 = 89406 5 = 0
Accepted triggers: 46047
Accepted triggers by helicity: (-) = 0 (+) = 45348
Charge Monitors (Micro Coulombs)
Upstream BCM gain x1 26801.33 x3 26891.27 x10 23065.28
Downstream BCM gain x1 28169.56 x3 57524.67 x10 24516.56
FIGURE 1