    User name jpchen

    Log entry time 23:50:05 on January20,2008

    Entry number 226443

    keyword=target cool-down: switched to cold return.

    Started cool-down at 8 pm. Just switched to cold return. We are drawing 8 g/s. Hall C is drawing typically 15 g/s. It should be confortablly below 25 g/s.
    Fan is off. I'll go into the hall to turn it back on (while reset
    Left Q3).
    The CERNOX temperature readings with fan off are:
    24.41/22.25/33.01/47.30. Pressure is at 36 psi (still chnaging).


    FIGURE 1