    User name V. Sulkosky

    Log entry time 06:50:42 on January28,2008

    Entry number 227677

    keyword=Cavity Monitor Current Linearity

    In Figure 1, the cavity monitor intensity ADC signals are plotted
    versus beam current (average from EPICS in datastream) for runs
    taken on January 27 (3394, 3397, 3414, 3473, 3478, 3480, 3481, 3482
    and 3487) with currents ranging from 5 uA to 80 uA. At 80 uA. the
    signals are close the saturation point.

    At the next chance, I should adjust the attenuation a bit to avoid
    saturation at 100 uA.


    FIGURE 1