User name fansler
Log entry time 11:57:27 on February 05, 2008
Entry number 228550
keyword=LQ2 Liquid He
I discover two bad isolation modules in the left Q2 he readback circuit.
I replaced both isolation modules. The magnet appears to have been
overfilling. Cryo reduced the control points to force the magnet liquid
level to fall below 100%. We are waiting for the liquid level to drop.
I will continue to monitor the circuit for evidence of additional
e4quipment failures.