User name V. Sulkosky
Log entry time 20:28:43 on March 21, 2008
Entry number 228717
This entry is a followup to: 228697
keyword=R-HRS Gas Cerenkov and T2
I have completed the cabling for the gas Cerenkov "or" of 10 PMTs. This
signal is plugged into channel # 11 of the LeCroy 4413 discriminator in
Crate # 2, slot 6.
In preparation for forming the T2 efficiency trigger, I have sent a copy
of this signal from the LeCroy 4518 delay module used for the gas
Cernknov in slot 7 from output # 3, channel # 11 into another delay
module in slot 11, channel # 15.
The following table shows
the input signals to the 4518 delay module:
LeCroy 4518 Delay Module
Slot # 11
Input Signal Input Channel Input Channel to MLU
-------------------------------------------------------------
"or" of S1 paddles 14 0
"or" of S2 paddles 15 1
"or" of Cer 10 PMTs 16 2
T1 trigger 1 ?
There seem to be multiple inputs into the delay module, which do not
seem to go anywhere. However, I was reluctant to disconnect them, since
I
wasn't sure about their purpose.
I also verified the location of the TDC and ADC for the gas Cerenkov.
Currently the HVs for S1, S2 and the gas Cerenkov are on in the R-HRS.
A copy of this log entry has been emailed to: camsonne@jlab.org,rom@jlab.org