    User name dhamilton

    Log entry time 05:43:10 on April 02, 2008

    Entry number 229495

    keyword=BB E/dE gain and discriminator threshold calibration with QE C data

    An iterative calibration of both HV settings and discriminator thresholds for the BB E and dE planes has been completed with Quasi-elastic data from a C foil target at 5 microamps.

    HV tune was already pretty good. Only small (5 - 15 V) changes were needed. The L-R balance was maintained throughout the calibration procedure.

    Discriminator threshold for the E plans has been set to 320 mV. The dE threshold is 200mV.

    Attached picture shows E vs dE plots for all paddles following calibration.

    Next step is to look at elastic data from the LH2 target and set minimum E threshold so as to accept 700 MeV/c protons.




    Figure 1