Main INDEX
Monthly INDEX
PREV
NEXT
User name dhamilton
Log entry time 08:06:07 on April 03, 2008
Entry number 229725
keyword=Owl Shift Summary
BigBite magnet switched on in preparation for elastic data-taking.
Coil current at 390A.
00:24 - elastic data taking at 3 microamps to check thresholds on BB E
and dE planes. Start run 2205.
00:45 - End of run 2205.
Target changed to carbon single foil.
Start of BB MWDC QE HV/threshold scan.
00:58 - Start run 2206 (HV 1350V, threshold 5V).
Run 2207 (HV 1350V, threshold 7.5V).
01:22 - problems with ROC5. No data in TDCs for E and dE paddles.
Tried to cycle the power on the crate. The problem is not solved.
Runs 2206 - 2225 are junk.
03:05 - Bryan Moffit arrived to help with the ROC5 problem.
03:45 - Hall status changed to controlled access. Bryan has went to the
hall to attempt to diagnose the problem with ROC5 TDCs.
04:30 - Bryan identified and fixed the problem (NIM level translator
failed, meaning that the ROC5 TDCs were not getting triggers). Hall
status being changed to Beam Permit.
04:45 - Requested 5 microamps for MWDC calibration runs.
Runs 2226 - 2229 (HV 1350V, thresholds 5, 7.5, 10, 12.5V).
Runs 2230 - 2233 (HV 1400V, thresholds 5, 7.5, 10, 12.5V).
Runs 2234 - 2237 (HV 1450V, thresholds 5, 7.5, 10, 12.5V).
05:55 - beam lost due to accelerator problems.
06:45 - beam restored to hall. Continue MWDC calibration.
Runs 2238 - 2241 (HV 1500V, thresholds 5, 7.5, 10, 12.5V).
Runs 2242 - 2245 (HV 1550V, thresholds 5, 7.5, 10, 12.5V).
07:40 - BB MWDC calibration runs complete.
Move target to 6 cm LH2. Turn MWDC HV off. Set prescales for coincidence
running.
BB coil current = 390A
Start run 2246.
Shift ended during run 2246.