• Main INDEX
  • Monthly INDEX
  • PREV
  • NEXT

    User name E.Chudakov

    Log entry time 23:32:55 on April 11, 2008

    Entry number 231453

    keyword=Moller results

    Results

    Beam polarization: -66.1+/-0.1(stat)+/-1.7%, measured on 4 targets, chi2=60/(11 ndf)

    Details

    Both the ADC and VtoF readouts for the target position encoder do not work. The EPICS variable is changing properly.

    We used 4 targets. The target #5 was measured in 2 points. The target #3 was measured at 2 different currents.

    HV: -1715.0 -1880.0 -2140.0 -2010.0 -1790.0 -1760.0 -1865.0 -1960.0 -1900.0 -1850.0

    Results

    List of measurements
    Target Encoders (EPICS) Beam Runs Target Beam
    X Y μA P P χ²
    5 1.040 1.387 0.55μA 12705-12706 7.97% -66.2+/-0.3% 0.5 / 1
    5 1.020 1.387 0.55μA 12707-12708 7.97% -66.9+/-0.3% 0.5 / 1
    4 1.660 1.387 0.55μA 12709-12710 7.44% -68.2+/-0.3% 0.3 / 1
    3 2.310 1.387 0.55μA 12711-12712 7.62% -65.8+/-0.3% 0.3 / 1
    3 2.310 1.387 0.30μA 12713-12714 7.62% -65.8+/-0.3% 18 / 1
    2 2.930 1.387 0.30μA 12715-12716 8.21% -66.1+/-0.3% 2 / 1
    average 12705-12716 -66.1+/-0.1% 60 / 11


    A copy of this log entry has been emailed to: gorbenko@jlab.org, glamazdi@jlab.org, romanip@jlab.org



    Figure 1