User name A. Camsonne
Log entry time 11:00:44 on April12,2008
Entry number 231580
This entry is a followup to: 231574
keyword=Left S2 paddle seems OK now
Looking at the plots, it seemed the EDTM was affected by the problem too. I thought it might be a timing issue. I asked to Bryan to redownload the trigger. It seems to have fixed the problem. I noticed the CAMAC crates are running a bit hot, Jack will have a look at them.
The empty part in the plot as a function of event number is when I unplugged the signal from ADC so it was replugged correctly.
FIGURE 1
FIGURE 2