Main INDEX
Monthly INDEX
PREV
NEXT
User name Yi Qiang
Log entry time 20:42:33 on April 22, 2008
Entry number 233820
This entry is a followup to: 233815
keyword=A detailed description of target after the power trip
After the power trip, the High Power Heater of loop1 went to 600W (its
maximum value) and the target temperature and target vacuum increased
repidly. The fan speed in Loop1 also went wrong. So we turned off the
HPH, moved the cryo target away from beam and changed the fan speed back
to 30 Hz. After a while, as the target temperature and pressure got
stabilized, we tried to turn on the HPH again and now it looked normal.
Since nothing is abnormal now, after we finish the Al dummy test, we will
continue to take production data on LH2.