Main INDEX
Monthly INDEX
PREV
NEXT
User name R. Michaels
Log entry time 14:04:17 on May 12, 2008
Entry number 237889
This entry is a followup to: 237880
keyword=T4 redefined. trigsetup defaults
The trigger T4 is now defined as the "or" of S1,S2 on L-HRS.
Normally it is a 2-out-of-3 trigger for efficiency measurement, but
there is no 3rd detector now. T4 will be ~90% junk (soft photons).
Recommend to keep 5 Hz of it anyway.
Also: Since downloading the trigger calls trigsetup, I tried to make
sure the trigsetup defaults are something reasonable. Default args
are 84 43 at present. Also clarified print statements in setbbite.
A copy of this log entry has been emailed to: moffit@jlab.org