• Main INDEX
  • Monthly INDEX
  • PREV
  • NEXT

    User name V. Sulkosky

    Log entry time 09:54:46 on August 12, 2008

    Entry number 242023

    keyword=S2m and A1 HV Problems Addressed

    There were two problematic HV channels one each for the S2m and A1 detectors that have been addressed. For the A1 detector, the 1461P card in slot # 14 has a problem with channel 2. Youcai Wang and I found a few spare 1461P cards in the Physics storage building. We swapped out the card in slot # 14 with one of these cards. The "new" card appears to work fine.

    When the HV was turned on for the sixth channel on the left side of S2m, the current readback for the HV remained at zero. We traced this to a bad HV cable. I suspect one or both of the connectors has a problem. We replaced the 'L6' HV cable with a new one. The current readback is now similar to the other S2m channels.

    A copy of this log entry has been emailed to: ycwang@jlab.org,segal@jlab.org,jiang@jlab.org