    User name V. Sulkosky

    Log entry time 22:54:37 on October 23, 2008

    Entry number 244485

    keyword=Runs 2094-2098 with T8 only and Deadtime

    Runs 2094 to 2098 are with the T8 pulser with different CODA configurations.
    It seems we have a lot of deadtime when running in the "Twoarm_RICH" configuration.
    I had pedestal suppression for ROC 3 and 4 turned off
    for these runs. I don't think that affects ROC 14 or 15.

    Here is a brief summary of what I saw:
    Run    Configuration       Trigger       Rate      Deadtime
    ------------------------------------------------------------
    2094    Twoarm_RICH         T8           500 Hz       88%
    2095    Twoarm_FB           T8           500 Hz        0%
    2096    Twoarm_FB           T8          1000 Hz        0%
    2097    Twoarm_RICH         T8          1000 Hz       95%
    2098    Twoarm_RICH         T8           100 Hz       47%
    