Log entry time 23:11:58 on October 26, 2008
Entry number 244818
Followups:
keyword=T1 and T6 rate difference
Triggers T1 and T6 are supposed to be copies of the same signals. I
noticed that when we have beam(run 2215), T1 rate is 10 times that of T6.
The thresholds for both the triggers were set to -80mV during this run.
This was not the case when I last checked with cosmics - both triggers
had equal rate when set to same threshold.
Note: Run 2215 was taken without any target and BB magnet turned on(710A).