    keyword=Badly damaged RTDs from the old reference cell

    Yi noted that the RTDs from the old reference cell were badly damaged. One is basically destroyed.

    We believe this was due to beam being slewed around on the reference cell during the ion chamber calibration procedure last night. I'll mention it to the PD tomorrow morning so the (new) MCC procedure can be amended and the crew chiefs told to strictly avoid any deliberate mis-steering with any glass cell in beam position.