NEXT
Make New Entry,
Make Followup Entry
User name jinhuang
Log entry time 02:16:03 on November 03, 2008
Entry number 246347
This entry is a followup to: 246333
keyword=Some discuss on Polarization loss during last Rotation-NMR combination
During target preparation at 23:04 yesterday, the target
apparent polarization dropped from 38%->31%. Related Logs are:
The possible reason of dropping are:
Field Rotation from Trans+ -> Long + @ Empty Target
Location
It's possible, bug not confirmed.
It also has been observed on Oct 28 that, there is an apparent
half Pol. loss on
pumping chamber during field rotation-NMR combination:
- NMR at Vertical Holding Field:
http://www.jlab.org/~bbsoftw/TargetReports/GenericLog/NMRField_befiore_p
ickup_28-10_17-24-17.html
- Rotation to Long Holding Field
- NMR at Long Holding Field :
http://www.jlab.org/~bbsoftw/TargetReports/GenericLog/NMRField_pickup_28
-10_17-31-51.html
- Rotation back to Vertical Holding Field
- Same NMR as step 1 :
http://www.jlab.org/~bbsoftw/TargetReports/GenericLog/NMRField_after_pic
kup_28-10_17-36-39.html
NMRs
If there is any problem with NMR, it should be large AFP loss,
because:
- For the first and second Junk NMR, the sweep rang is
24.982G->24.982G. So there
shouldn't be much phisical affect on target
- The thrid junk NMR and 4th NMR (correct one) should send
exactly same RF& field
sweep to target. The "junk point" is just that after taking
data for the 3rd NMR,
the way software processing data (background subtraction)
is not correctly setted.
- Both 3rd and 4th NMR show similar 31% low polarization.
- The 4th NMR is using 1mV rather than 300mV as sensitivity
of lockin amplifier. However,
it has been tested that, as long as there is no overflow,
the calibration constance
should preserve.
- the holding field is slightly different than that when
taking beam
- AFP loss at pickup coil is uncertain now. (tested to be ~1%
on Transverse Pumping
@ BeO location)
Calibration error
The calibration chain seems fine :
Vertical
EPR -> NMR @ Optics Position, Vertical Holding Field,
Long RF, Trans Pickup
->
NMR @ PickUp Position, Long Holding Field, Vert RF, target
chamber Pickup ->Vertical
EPR
The ending EPR shows to be consist within 10% relative
error. The other chain
is
Trans
EPR -> NMR @ Empty Position, Trans Holding Field, Long
RF, Vert Pickup
It's suggested that next field rotation been taken slow
A copy of this log entry has been emailed to: jpchen@jlab.org , yqiang@jlab.org