    User name Jin Huang

    Log entry time 03:36:02 on November 04, 2008

    Entry number 246567

    This entry is a followup to: 246566

    Followups:

    keyword=Target Long -> Trans

    
    He3 cell is left in beam taking longitudinal data until 33.6% Pol. Then 
    
    1. an NMR @ Pickup Location (Pol = 33.6%)
       https://www.jlab.org/~bbsoftw/TargetReports/GenericLog/NMRField_04-
    11_02-25-29.html
    
    2. rotate to trans+ direction
    
    3. rise to optics location, NMR (cross calib, Pol. = 30.1%)
       https://www.jlab.org/~bbsoftw/TargetReports/GenericLog/NMRField_04-
    11_02-43-14.html
    
    4. lowered to beam location, NMR calibration 
       (Pol. = 2.13 %/mV x (X Channel Average Sweep Up/Down)
       https://www.jlab.org/~bbsoftw/TargetReports/GenericLog/NMRField_04-
    11_02-46-07.html
    
    5. Wait 30min, NMR shows (Pol. = 36.8%)
       https://www.jlab.org/~bbsoftw/TargetReports/GenericLog/NMRField_04-
    11_03-16-53.html
    
    
    ---------------------------------
          Calibration Chain
    ---------------------------------
    
    
    Using Calib of 47.65% Pol @ RF=long. , Pickup = Vertical, Target = 
    Optics
    
    X Channel Sweep Down = 22.0mV (Sweep up has distortion)
    
    http://www.jlab.org/~adaq/halog/html/0811_archive/081103141754.html
    
    ----------------------------------
    
    NMR Just after Pumping :
    
    Target Position : Optics
    RF Position : Long
    Holding Field : Trans
    Pumping Chamber Pick Up Coil : Vert
    
    X Channel Sweep Down = 13.9mV
    
    => Pol = 30.1%
    
    https://www.jlab.org/~bbsoftw/TargetReports/GenericLog/NMRField_04-
    11_02-43-14.html
    
    ----------------------------------
    
    Another immediate follow up NMR @ He3 Cell Location
    
    Target Position : He3 Cell
    RF Position : Long
    Holding Field : Trans
    Pumping Chamber Pick Up Coil : Vert
    
    X Channel Average Sweep Up/Down = 14.1mV
    
    https://www.jlab.org/~bbsoftw/TargetReports/GenericLog/NMRField_04-
    11_02-46-07.html
    
    ----------------------------------
    
    So calibration constant for 
    He3 Cell Position
    Trans Holding Field
    Long RF
    Vert Pickup 
    
    Pol. = 2.13 %/mV x (X Channel Average Sweep Up/Down)