    User name V. Sulkosky

    Log entry time 11:00:22 on November 05, 2008

    Entry number 246861

    keyword=BPM Calibration Completed

    I have finally finished the analysis of the BPM calibration. The process
    ended up being a lot more involved due to the quality of the data.

    Figure 1 shows the BPM A and B x and y positions calibrated against the
    Harp scans from the bulls eye scan taken last Wednesday. I have updated
    all BPM and raster databases for the online analysis and for spot++.

    Figure 2 shows the beam positions for run 2853 with the obsolete
    calibration constants. Figure 3 shows the beam positions with the new
    coefficients.

    For this calibration, I did not use the Harp A X position, since there was
    a 1 mm offset compared to all the previous data taken last year. Instead
    I used my fit of the Harp A x position versus EPICS BPM A x position, which
    I believe to be a more reliable number until the harps are resurveyed.

    Here are the new coefficients:
    BPMA constants:
    -0.738959 0.759596 0.758844 0.747731 -0.000311576 0.00109736
    BPMB constants:
    -0.686875 0.737689 0.735834 0.691631 -0.00100634 0.00228915



    A copy of this log entry has been emailed to: brads,jiang,rom



    Figure 1



    Figure 2



    Figure 3