    User name Jin

    Log entry time 04:02:39 on November 14, 2008

    Entry number 248715

    keyword=Target Activity of Nov-14 Owl

    There was a successful NMR cross calibration. 
    
    However, before next test, the correction coil current is not correctly 
    setup. And polarization goes to low (~10%). So an spin up has been 
    started for 
            Spin Down. 
            NMR Setting same as Spin Flip
            Time Interval = 120 min
            Oven = 230C
            Laser = 3 Comet Laser:
               1.Rutgers@34A.20C
               2.WM@35A.26C
               3.JLab@36A.25C
    
    
    Start Point:
    			Pick up Coil Pos
          Trans - Holding Field
          Spin Antiparallel to Field
          Pol.=~60%
    11:06 NMR Freq Sweep -> Pol = ~60% (42mV)
              RTD#0->6=67C 133C 167C 106C 52C 223C 221C 
    11:20 Bigbite ->710A
    11:51 Pumping Laser Off
          From now on there is no good NMR pol first guess 
          since cell temperature changes 
    12:33 Temp Stable to 
              Oven=330C
              Coil=-4.601A,6.232A,0.712A
              RTD#0->6=66C 134C 167C 107C 53C 224C 221C
    12:40 Target -> He3
    12:43 Freq NMR @ He3 
          -> Pol=~60% (40mV), Peak @ 81kHz 
    12:53 Temp Stable to 
              Oven=330C
              Coil=-4.599A,6.230A,0.712A
              RTD#0->6=40C 71C 140C 61C 38C 223C 220C
    --------------------------------------------------------
                  NMR Cross Calibration Start
    --------------------------------------------------------
    12:54 Freq NMR @ He3 
          -> Pol=50-60% 33mV 
    01:00 Field -> Long +
    01:02 Target -> Pick Coil
    01:13 Temp Stable to 
              RTD#0->6=64c 130c 167c 105c 52c 224C 221C
    01:14 Field NMR @ Ref 
          -> Pol=~56% (calib const changes when he3 temperature changes)
    01:15 Target -> He3
    01:27 Temp Stable to 
              RTD#0->6=40C 71C 137C 62C 38C 223C 221C
    01:27 Freq NMR @ He3 
          -> Pol=50-60% (31mV)
    --------------------------------------------------------
                 NMR Cross Calibration Ends
    --------------------------------------------------------
    01:50 Temperature change screen shot
    02:06 NMR Field Sweep. Pol.=~40~50%
    02:09 NMR AFP loss test with 2 field sweep 
          -> AFP Loss =~ 0.7%
    02:25 NMR AFP loss test with 5 Frequency sweep (Same Setting As Spin 
    Flip) 
          -> AFP Loss =~ 0.7%
    2:38  Correction Coil On to 
          Large = 5.00A
          Small = 5.00A
    2:39  NMR AFP loss test with 2 Frequency sweep with Correction Coil ON 
          -> AFP Loss = ~20%, 
          -> Field Gradiant = Extream High, Peak Width is 5 time of normal
    2:47  Correction Coil On to 
          Large = 1.00A
          Small = 1.00A
    2:47  NMR AFP loss test with 5 Frequency sweep with Correction Coil ON 
          -> Field Gradiant = High, Peak Width is 1.5 time of normal
          -> AFP Loss = ~ 2.7%
    2:58  Correction Coil On to 
          Large = 0.50A
          Small = 0.50A 
    2:59  NMR AFP loss test with 5 Frequency sweep with Correction Coil ON 
          -> AFP Loss = ~ 1.5%
    3:10  Correction Coil OFF
    3:12  NMR AFP loss test with 5 Frequency sweep with Correction Coil ON 
          -> AFP Loss = ~ 1.0%
          -> Pol. = ~10%
    3:30  Laser ON, Wait Temperature Stablized
    3:47  Start Spin Up
            Spin Down. 
            NMR Setting same as Spin Flip
            Time Interval = 120 min
            Oven = 230C
            Laser = 3 Comet Laser:
               1.Rutgers@34A.20C
               2.WM@35A.26C
               3.JLab@36A.25C
    
    
    


    A copy of this log entry has been emailed to: yqiang, jpchen



    Temperature change during cross calibration