• Main INDEX
  • Monthly INDEX
  • PREV
  • NEXT
    Make New Entry, Make Followup Entry

    User name Moffit

    Log entry time 12:01:59 on November 18, 2008

    Entry number 249540

    keyword=HAPPEX DAQ - Moved x10 BCM to adc15_1

    x10 BCM (bcm4) signal was saturating adc22_3. So I moved it to adc15_1. Signal at 12uA sits around 30k ADC channels. I suspect they will run up to 15uA... so we should be good from here on out.

    This occurred during run 30419. The database was also changed but will not take into effect until the next run (30420) is started.

    After the pedestal is calibrated, we'll use this bcm for charge asymmetry feedback.

    A copy of this log entry has been emailed to: rom,mcnulty,mercado